ARTEL Extreme Pipetting Expedition
Mission 1 Results
Mission 1 results
Mission 2 results
Mission 3 results
Mission 4 results
RESULTS
MISSION 2 – YELLOWSTONE NATIONAL PARK
The Effects of Pipetting Warm and Cold Fluids on Pipette Performance

The volumes of aqueous samples, which were not in thermal equilibrium with the pipettes and tips used to deliver them, exhibit significant deviation compared to volumes pipetted using thermally equilibrated samples, pipettes and tips. The results obtained for delivering samples, which were in ambient temperature equilibrium with the tips and pipettes used for delivery, served as reference volumes to which the results obtained for warm or cold samples were compared.

Pipetting cold samples (1 – 4°C) resulted in over-delivery of up to 37%, while warm liquids (45°C) were under-delivered by up to 23%. Pipetting errors increased as the set volume of the pipette decreased. When pipettes were used at their minimum specified volume range, significantly larger errors were observed as compared to using them at their nominal volume setting.

Figure 1 shows the errors induced by pipetting warm and cold liquids with a 2 μL pipette, set at its nominal volume (right graph) and at 10% of its nominal volume (0.2 μL, left graph).

Figure 2 summarizes the results obtained for a 20 μL (nominal) pipette. It is obvious that the induced error is larger when the pipette is used at its smallest volume setting, as compared to its nominal set volume. The observed errors are smaller for this larger volume pipette compared to the observed errors for the tested 2 μL pipette (Figure 1).
Results: Mission 2
Figure 1. Pipetting of warm and cold samples with a 2 μL pipette, set at 0.2 μL (left graph), and at 2.0 μL (right graph).
Results: Mission 2
Figure 2. Pipetting of cold samples with a 20 μL pipette, shows a larger error when set at 2.0 μL (left graph) than at its nominal volume of 20.0 μL (right graph). The magnitude of the induced error is smaller than for the 2 μL pipette (Figure 1).
Extreme weather conditions encountered in Yellowstone National Park during this mission of the Extreme Pipetting Expedition made it impossible to complete data collection of data points for all pipettes at all temperatures.

The significance of this field data, collected in an environment emblematic for real-life laboratory conditions, warranted proper scientific investigation of the errors resulting from pipetting liquids which are warmer or colder than the used pipettes and tips.

The evaluation of three commonly used variable volume pipettes (20 μL, 200 μL, and 1000 μL) from three leading manufacturers is shown in Figures 3, 4, and 5. The same test protocol as in the field was followed, but under carefully controlled laboratory conditions. As pipetting heated samples at 60°C is not uncommon in many procedures, pipette testing at this temperature point was added to the protocol.  

It is clearly visible in Figure 3 that all of the pipettes, at each tested volume setting, over-delivered cold samples, which were thermostatted at 4°C.

Warm samples, in contrast, were consistently under-delivered as shown in Figure 4 (samples at 37°C) and Figure 5 (samples at 60°C).

All of the above results also indicate that the error associated with pipetting warm or cold liquids is significantly larger for pipettes designed to deliver small volumes. It is furthermore obvious, that errors increase when the pipettes were used at or close to their minimum specified volume ranges, as compared to their nominal volumes.

These laboratory results are in good agreement with the field results from Yellowstone National Park.

THE CONCLUSION
Researchers who are pipetting warm or cold liquids need to be aware that this technique is prone to introduce significant errors into common laboratory procedures. Cold liquids tend to be delivered in excess quantity, while warm liquids tend to be under-delivered. Depending on pipette manufacturer, volume set point and temperature of the sample, these errors can exceed 65%, with small volumes to be impacted most.

Whenever possible, it is recommended to transfer liquids that are equilibrated to room temperature. When using protocols necessitating the handling of cold or warm liquids with an air displacement pipette, it is recommended that the researcher determines the pipette inaccuracy of the used pipette/tip/temperature combination prior to the experiment. Since it is not always feasible to determine the precise aberration from the calibrated volume at any given temperature, volume, and tip combination, everyone interpreting the data should be aware of the potentially very significant error introduced by pipetting warm and cold liquids.
Results: Mission 2
Figure 3. Pipetting liquids thermostatted at 4°C with pipettes set at or close to the specified minimum and nominal volumes, results in over-delivery of the samples.
Results: Mission 2
Figure 4. Under-delivery of samples thermostatted at 37°C.
Results: Mission 2
Figure 5. Errors associated with pipetting warm samples increase with increasing sample temperature, as shown here for 60°C warm samples.

© Artel, Inc.  All Rights Reserved.