ARTEL Study Shows that Pipetting Warm and Cold Liquids
Causes Error of up to 37 Percent
Visits Yellowstone for Mission #2 of the Extreme Pipetting Expedition
Westbrook, ME (July 17, 2007) – Data taken at Yellowstone National Park during Mission #2 of the ARTEL Extreme
Pipetting Expedition show that pipettes deliver volumes inaccurate by up to 37
percent when handling fluids at temperatures different from the pipettes.
Called thermal disequilibrium, this source of error is prevalent in today’s laboratories, as many assays and tests require the addition of liquids, which
are warmer or colder than the laboratory environment and the actual pipettes.
Yellowstone was selected for Mission #2 of the Expedition because it is
emblematic of thermal variation and disequilibrium. The Extreme Pipetting
Expedition is a multi-phase, year-long scientific study to investigate and
illustrate the impact of laboratory conditions on data integrity.
ARTEL is releasing the results from Mission #2 of the Extreme Pipetting
Expedition at the 2007 AACC Annual Meeting & Clinical Lab Expo, San Diego, CA, July 17-19, Booth #4143. The data will also
be published at www.artel-usa.com/extreme.
“From our testing at Yellowstone, it is clear that pipettes show a bias in volume
delivery when dispensing fluids at different temperatures, and the error is
especially significant when handling small liquid quantities,” says George Rodrigues, Ph.D., Senior Scientific Manager at ARTEL and Expedition
Leader. “Laboratories must account for this variation to ensure good accuracy and
precision in test results.”
Data taken at Yellowstone show that error caused by thermal disequilibrium was
most significant when using smaller volume pipettes at their minimum volumes.
For example, a two-microliter variable-volume pipette set to deliver 0.2
microliters over-delivered cold liquid (3°C) by 37 percent and under-delivered warm liquid (45°C) by 24 percent, compared with liquid at room temperature (21°C). When set to deliver its maximum volume (two microliters), the pipette
over-delivered cold liquid by one percent and under-delivered warm liquid by
seven percent.
As expected, error was present, but smaller, when working with larger liquid
volumes. The 20-microliter pipette set to deliver two microliters
over-delivered cold liquid by four percent. When the pipette was set to deliver
its maximum volume of 20 microliters, the over-delivery amounted to 0.6
percent.
“There are not many laboratory managers who would knowingly accept error of more
than a few percent, let alone error of 37 percent,” says Doreen Rumery, Quality Control Manager at ARTEL and Extreme Pipetting
Expedition member. “Yet many technicians frequently pipette warm and cold liquids in critical
testing and assay work without accounting for volume variation.”
The aim of the Extreme Pipetting Expedition is to raise awareness about common
laboratory extremes that can affect data quality, including the thermal
disequilibrium phenomenon. Restriction enzymes used in nucleic acid work, for
example, are frequently handled at ice temperature (0°C), and higher temperatures are encountered when handling mammalian cell
cultures (37°C) or polymerase chain reaction (PCR) solutions (60°C or higher).
The resulting volume differences can be explained by the air displacement
operation of pipettes. The dead air volume in pipettes acts to pull liquid into
the pipette tip. When inserting the pipette tip into warm fluid, the air inside
the tip is at ambient temperature. During aspiration, the pipette tip heats up,
causing the air to expand and push liquid out of the tip. This causes less
liquid to be dispensed, leading to variation between the target and delivered
volumes. The opposite occurs when pipetting cold liquids.
“Since the magnitude of error is dependent on a number of protocol-specific
details, such as pipetting speed and type of sample container, a
one-size-fits-all correction factor does not currently exist,” says Rodrigues. “ARTEL will continue to study thermal disequilibrium to help laboratories develop
strategies to minimize this source of error.”
The naturally unstable conditions at Yellowstone exemplified the challenges
laboratories face with sometimes unpredictable or uncharacterized changes in
the testing environment. For example, humidity and ambient air temperature greatly fluctuated through
stages of the testing, leading to volume variation and, in some cases,
inability to isolate fluid temperature as the source of error. Affected data
were not reported. To reduce environmental variation, ARTEL repeated the
testing in a controlled laboratory and summarized the results in a poster
entitled “Errors Associated with Pipetting Warm and Cold Liquids,” which is available for download at www.artel-usa.com/extreme.
Volumes were measured using the ARTEL PCS® (Pipette Calibration System), a portable, robust measurement technology based
on ratiometric photometry. The system is unaffected by the environment and is
highly accurate and precise, even at small volumes.
# # #
ARTEL is the worldwide leader in liquid handling quality assurance. ARTEL manufactures the most accurate, precise and easy-to-use systems for
ensuring data integrity in liquid volume measurement. ARTEL systems, based on
proprietary Ratiometric Photometry, conform to ISO 8655-7 and provide
NIST-traceable results. In addition, ARTEL provides liquid handling quality
assurance support and consultation services including on-site pipetting
technique training and certification for pipette users and laboratory managers,
as well as Liquid Handler Performance Verification services for automated
laboratories. Since 1982, ARTEL technology has been proven in daily use in
thousands of laboratories including pharmaceutical, clinical, forensic, public
health and environmental. Leading institutions such as Amgen, ARUP,
Bristol-Myers Squibb, Celgene, Centers for Disease Control, DuPont, Eli Lilly,
FBI, Genentech, Genomic Health, Mayo Clinic, Memorial Sloan-Kettering Cancer
Center, Merck, Pfizer, Procter & Gamble and U.S. FDA depend on ARTEL technology to help them meet their
objectives for quality and productivity.
|
|
||||
|
|
|||||
|
|
|||||
|
|
|
|
|
|
|